96062 钳式探头
φ24 mm AC 100 A, 负载电流测量。
型号 | 96062 | |
Measureable conductor size | Max. φ24 mm | |
Measuring range | AC 0 to 100 Arms (141 A peak) | |
Output voltage | AC 0 to 500 mVrms (5 mV/A) | |
Accuracy (sine waveinput) | Band width | ±0.5%rdg±0.1 mV( 50/60 Hz) ±1.0%rdg±0.2 mV (40 Hz to 1 kHz) |
Phase | Within ±2.0 deg (at 1 A to 100 A, 45 Hz to 65 Hz) | |
Temperature & humidity range (guaranteed accuracy) | 23 ±5°C, 85%RH or less(no ondensation) | |
Rated voltage between circuit and ground | AC 300 Vrms Max. | |
Operating temperature & humidity range | 0 to 50°C, 85%RH or less (no ondensation) | |
Storage temperature & humidity range | -20 to 60°C, 90%RH or less (no ondensation) | |
Temperature coefficient | In the range of 0 to 50°C and the amount of change from 23°C is ±0.5% rdg ± 0.1 mV or less | |
Maximum allowable input | AC 100 Arms continuous (50/60Hz) | |
Influence of the conductor location | Within ±0.5% rdg | |
Influence of the external magnetic field | 0.2 A equivalence or less (400 A/m, 50/60Hz) | |
Output impedance | Approx. 11 Ω | |
Withstand voltage | AC 3.51 kVrms / 5 s Between jaw and case, case and output terminal, output terminal and jaw. | |
Environmental condition | Altitude up to 2000 m, indoor use | |
Recommended calibration period | 1 year | |
Dimension | 60(W)×100(H)×26(D) mm (excluding projections) | |
Cable length | Approx.3m | |
Output terminal | MINI DIN 6PIN | |
Weight | Approx.160g | |
Accessory | Instruction manual, cable marker | |
Safety standards | Compliant standards: EN 61010-1, EN IEC 61010-2-032 Measurement category III (rated voltage between circuit and ground:300Vrms) Pollution degree 21 | |
EMC | Compliant standards: EN 61326-1 Class A, EMC standards of Australia and New Zealand EN55011 Class A, Group 12 Korea Electromagnetic Conformity Standard ( 한국 전자파적합성기준 ) This product is a Class A (for industrial environment) product. Operation of this product in a residential area may cause radio interference in which case the user will be required to correct the interference. Immunity Influence (Performance degradation with self reset) | |
Environmental standards3 | EU RoHS Directive compliant |